pressure < -10cm h2o warning |
sub-atmospheric pressure alarm |
|
defaults mv < 3LPM |
volume monitor |
|
exhaled tv= |
tv + fgf - circuit compliance in machines without fgd. |
|
what 3 alarms are displayed in the volume monitor |
1. apnea-volume alarm |
|
digital display of tv, bpm and mv |
volume monitor |
|
caution on apnea vol alarm |
breath (>80ml) not detected for 15sec |
|
warning on apnea vol alarm |
breath (>80ml) no detected for 30sec. |
|
if mv < alarm setting |
low minute vol |
|
reversed flow >20ml through sensor |
reverse flow |
|
upper airways heat and humidify inhaled gases to |
37degc and 95% humidity |
|
cilia beat |
100-1500cycles/min |
|
mucociliary transport works best at |
80-90%RH and 32-33deg C |
|
actual mass of water in a given vol of gas at a given temp, mg/L |
absolute humidity |
|
max mass of water vapor that a given vol of gas can hold at a given temp |
maximum (saturated) humidity |
|
water content directly related to |
temp |